Configuring Serial (RS-232) Touch via VGA-to-Serial Y-Cable on Elo 03-Series Medical PCAP Monitors
Audience: Public / Elo Customers, Healthcare IT, Field Technicians & Integrators
Article Type: Hardware Setup & Serial Driver Configuration Guide
Select Elo 03-Series Medical Desktop Touchmonitors (TouchPro® PCAP) support legacy Serial (RS-232 / COM port) touch communication utilizing an optional custom VGA-to-Serial Y-cable assembly. This capability allows modern medical PCAP displays to interface with legacy medical equipment, diagnostic carts, and specialized operating systems (such as Windows XP) that mandate RS-232 serial touch controllers.
Compatible Medical Monitor Models
| Model | Description & Display Size |
|---|---|
| 1903LM | 19-inch Medical Desktop Touchmonitor (5:4 Standard Aspect Ratio) |
| 2203LM | 22-inch Medical Desktop Touchmonitor (16:9 Widescreen FHD) |
| 2403LM | 24-inch Medical Desktop Touchmonitor (16:9 Widescreen FHD) |
| 2703LM | 27-inch Medical Desktop Touchmonitor (16:9 Widescreen FHD) |
Required Hardware Accessory: Elo VGA-to-Serial Y-Cable
Serial touch functionality requires the specialized Elo custom Y-cable assembly (sold separately). Contact your authorized Elo Sales Representative or medical hardware distributor to purchase:
E810917Description:
CABLE, VGA TO DB9/VGA Y-CABLE, MEDICALReference Article: E810917 - VGA to Serial Y-Cable Specifications & Pinout
Touch Driver Download & Installation Steps
When operating via RS-232 serial connection or deploying on Windows XP, follow the driver installation procedure below:
- Visit the official Elo Touch Driver & Software Download Portal.
- Select your target host operating system from the Operating System dropdown menu.
- Select TouchPro (Projected Capacitive) or Serial Touch under the Technologies menu.
-
Download the driver package.
⚠️ Driver Version Requirement: For RS-232 serial touch operation on Windows, the installed Elo Multi-Touch driver version must be Version 7.0 or higher. - Run the installer, select the Serial / COM Port installation option during setup, select the corresponding COM port (e.g., COM1, COM2), and complete the calibration wizard.
⚠️ Operational Constraint for Serial Touch:
Due to baud rate and serial throughput bandwidth constraints inherent to RS-232 protocol specifications, touch tracking over serial COM connections is limited to a maximum of two (2) simultaneous touch points. Full 10-touch gesture tracking requires standard USB connection.